Open by-pass

Open by-pass fume hoods are designed for operation on constant air volume exhaust systems. The air by-pass provides for an alternate route for air to enter the hood as the sash is closed. The size of the by-pass is set so that, as the sash is closed, the velocity of the air increases to no more than three and one half times the velocity with the sash fully open. As a result, the static pressure loss through the front opening of these hoods is insignificant when compared to the pressure loss through the rear baffle and duct entry. Since the hood static pressure and the exhaust volume remain essentially constant, regardless of the sash position these hoods are classified as Constant Volume fume hoods.